Vulnerability Details : CVE-2023-1966
Instruments with Illumina Universal Copy Service v1.x and
v2.x contain an unnecessary privileges vulnerability. An unauthenticated
malicious actor could upload and execute code remotely at the operating system
level, which could allow an attacker to change settings, configurations,
software, or access sensitive data on the affected product.
Vulnerability category: Execute code
Products affected by CVE-2023-1966
- cpe:2.3:o:illumina:iscan_firmware:4.0.0:*:*:*:*:*:*:*
- cpe:2.3:o:illumina:iscan_firmware:4.0.5:*:*:*:*:*:*:*
- cpe:2.3:o:illumina:iseq_100_firmware:*:*:*:*:*:*:*:*
- cpe:2.3:o:illumina:miniseq_firmware:*:*:*:*:*:*:*:*
- cpe:2.3:o:illumina:miseq_firmware:*:*:*:*:*:*:*:*
- cpe:2.3:o:illumina:miseqdx_firmware:*:*:*:*:-:*:*:*
- cpe:2.3:o:illumina:miseqdx_firmware:4.0:*:*:*:ruo:*:*:*
- cpe:2.3:o:illumina:nextseq_500_firmware:4.0:*:*:*:*:*:*:*
- cpe:2.3:o:illumina:nextseq_550_firmware:4.0:*:*:*:*:*:*:*
- Illumina » Nextseq 550dx Firmware »Versions from including (>=) 1.0.0 and up to, including, (<=) 1.3.1cpe:2.3:o:illumina:nextseq_550dx_firmware:*:*:*:*:-:*:*:*
- cpe:2.3:o:illumina:nextseq_550dx_firmware:*:*:*:*:-:*:*:*
- cpe:2.3:o:illumina:nextseq_550dx_firmware:4.0:*:*:*:ruo:*:*:*
- cpe:2.3:o:illumina:nextseq_1000_firmware:1.4.1:*:*:*:*:*:*:*
- cpe:2.3:o:illumina:nextseq_2000_firmware:1.4.1:*:*:*:*:*:*:*
- cpe:2.3:o:illumina:novaseq_6000_firmware:*:*:*:*:*:*:*:*
- cpe:2.3:o:illumina:novaseq_6000_firmware:1.8:*:*:*:*:*:*:*
Exploit prediction scoring system (EPSS) score for CVE-2023-1966
0.16%
Probability of exploitation activity in the next 30 days
EPSS Score History
~ 34 %
Percentile, the proportion of vulnerabilities that are scored at or less
CVSS scores for CVE-2023-1966
Base Score | Base Severity | CVSS Vector | Exploitability Score | Impact Score | Score Source | First Seen |
---|---|---|---|---|---|---|
7.4
|
HIGH | CVSS:3.1/AV:P/AC:L/PR:L/UI:N/S:C/C:H/I:H/A:H |
0.7
|
6.0
|
ICS-CERT | |
9.8
|
CRITICAL | CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H |
3.9
|
5.9
|
NIST |
CWE ids for CVE-2023-1966
-
The product performs an operation at a privilege level that is higher than the minimum level required, which creates new weaknesses or amplifies the consequences of other weaknesses.Assigned by: ics-cert@hq.dhs.gov (Secondary)
-
The product does not properly assign, modify, track, or check privileges for an actor, creating an unintended sphere of control for that actor.Assigned by: nvd@nist.gov (Primary)
References for CVE-2023-1966
-
https://support.illumina.com/downloads/illumina-universal-copy-service-1-0.html
Illumina Universal Copy Service VulnerabilityVendor Advisory
-
https://www.cisa.gov/news-events/ics-medical-advisories/icsma-23-117-01
Illumina Universal Copy Service | CISAThird Party Advisory;US Government Resource
Jump to