An improper sanitization of input vulnerability in B. Braun SpaceCom2 prior to 012U000062 allows a remote unauthenticated attacker to gain user-level command-line access by passing a raw external string straight through to printf statements. The attacker is required to be on the same network as the device.
Published 2021-08-25 12:15:17
Updated 2022-07-12 17:42:04
Source MITRE
View at NVD,   CVE.org
Vulnerability category: Overflow

Products affected by CVE-2021-33886

Exploit prediction scoring system (EPSS) score for CVE-2021-33886

0.33%
Probability of exploitation activity in the next 30 days EPSS Score History
~ 70 %
Percentile, the proportion of vulnerabilities that are scored at or less

CVSS scores for CVE-2021-33886

Base Score Base Severity CVSS Vector Exploitability Score Impact Score Score Source First Seen
5.8
MEDIUM AV:A/AC:L/Au:N/C:P/I:P/A:P
6.5
6.4
NIST
8.1
HIGH CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:N
2.8
5.2
MITRE
8.8
HIGH CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
2.8
5.9
NIST

CWE ids for CVE-2021-33886

References for CVE-2021-33886

Jump to
This web site uses cookies for managing your session, storing preferences, website analytics and additional purposes described in our privacy policy.
By using this web site you are agreeing to CVEdetails.com terms of use!