Medtronic MyCareLink Smart 25000 all versions are vulnerable to a race condition in the MCL Smart Patient Reader software update system, which allows unsigned firmware to be uploaded and executed on the Patient Reader. If exploited an attacker could remotely execute code on the MCL Smart Patient Reader device, leading to control of the device.
Published 2020-12-14 20:15:13
Updated 2020-12-15 16:40:36
Source ICS-CERT
View at NVD,   CVE.org
Vulnerability category: Execute code

Exploit prediction scoring system (EPSS) score for CVE-2020-27252

0.31%
Probability of exploitation activity in the next 30 days EPSS Score History
~ 70 %
Percentile, the proportion of vulnerabilities that are scored at or less

CVSS scores for CVE-2020-27252

Base Score Base Severity CVSS Vector Exploitability Score Impact Score Score Source First Seen
9.3
HIGH AV:N/AC:M/Au:N/C:C/I:C/A:C
8.6
10.0
NIST
8.1
HIGH CVSS:3.1/AV:N/AC:H/PR:N/UI:N/S:U/C:H/I:H/A:H
2.2
5.9
NIST

CWE ids for CVE-2020-27252

  • The product checks the state of a resource before using that resource, but the resource's state can change between the check and the use in a way that invalidates the results of the check. This can cause the product to perform invalid actions when the resource is in an unexpected state.