An issue was discovered on Dr Trust ECG Pen 2.00.08 devices. Because the Bluetooth LE support is implemented without a requirement for pairing or security, any attacker can access the GATT server of the device and can sniff the data being broadcasted while a measurement is being done. Also, saved data can also be extracted over a Bluetooth connection. In addition, an attacker can launch a man-in-the-middle attack against data integrity.
Published 2020-08-26 16:15:13
Updated 2021-07-21 11:39:24
Source MITRE
View at NVD,   CVE.org

Products affected by CVE-2020-15486

Exploit prediction scoring system (EPSS) score for CVE-2020-15486

0.05%
Probability of exploitation activity in the next 30 days EPSS Score History
~ 20 %
Percentile, the proportion of vulnerabilities that are scored at or less

CVSS scores for CVE-2020-15486

Base Score Base Severity CVSS Vector Exploitability Score Impact Score Score Source First Seen
3.3
LOW AV:A/AC:L/Au:N/C:P/I:N/A:N
6.5
2.9
NIST
6.5
MEDIUM CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N
2.8
3.6
NIST

References for CVE-2020-15486