Vulnerability Details : CVE-2020-12048
Phoenix Hemodialysis Delivery System SW 3.36 and 3.40, The Phoenix Hemodialysis device does not support data-in-transit encryption (e.g., TLS/SSL) when transmitting treatment and prescription data on the network between the Phoenix system and the Exalis dialysis data management tool. An attacker with access to the network could observe sensitive treatment and prescription data sent between the Phoenix system and the Exalis tool.
Products affected by CVE-2020-12048
- cpe:2.3:o:baxter:phoenix_x36_firmware:3.36:*:*:*:*:*:*:*
- cpe:2.3:o:baxter:phoenix_x36_firmware:3.40:*:*:*:*:*:*:*
Exploit prediction scoring system (EPSS) score for CVE-2020-12048
0.09%
Probability of exploitation activity in the next 30 days
EPSS Score History
~ 38 %
Percentile, the proportion of vulnerabilities that are scored at or less
CVSS scores for CVE-2020-12048
Base Score | Base Severity | CVSS Vector | Exploitability Score | Impact Score | Score Source | First Seen |
---|---|---|---|---|---|---|
5.0
|
MEDIUM | AV:N/AC:L/Au:N/C:P/I:N/A:N |
10.0
|
2.9
|
NIST | |
7.5
|
HIGH | CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N |
3.9
|
3.6
|
NIST |
CWE ids for CVE-2020-12048
-
The product transmits sensitive or security-critical data in cleartext in a communication channel that can be sniffed by unauthorized actors.Assigned by:
- ics-cert@hq.dhs.gov (Secondary)
- nvd@nist.gov (Primary)
References for CVE-2020-12048
-
https://www.us-cert.gov/ics/advisories/icsma-20-170-03