Vulnerability Details : CVE-2020-12041
The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24) telnet Command-Line Interface, grants access to sensitive data stored on the WBM that permits temporary configuration changes to network settings of the WBM, and allows the WBM to be rebooted. Temporary configuration changes to network settings are removed upon reboot.
Products affected by CVE-2020-12041
- cpe:2.3:o:baxter:sigma_spectrum_infusion_system_firmware:8.0:*:*:*:*:*:*:*
Exploit prediction scoring system (EPSS) score for CVE-2020-12041
0.28%
Probability of exploitation activity in the next 30 days
EPSS Score History
~ 48 %
Percentile, the proportion of vulnerabilities that are scored at or less
CVSS scores for CVE-2020-12041
Base Score | Base Severity | CVSS Vector | Exploitability Score | Impact Score | Score Source | First Seen |
---|---|---|---|---|---|---|
7.5
|
HIGH | AV:N/AC:L/Au:N/C:P/I:P/A:P |
10.0
|
6.4
|
NIST | |
9.4
|
CRITICAL | CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:L/A:H |
3.9
|
5.5
|
NIST |
CWE ids for CVE-2020-12041
-
The product specifies permissions for a security-critical resource in a way that allows that resource to be read or modified by unintended actors.Assigned by:
- ics-cert@hq.dhs.gov (Secondary)
- nvd@nist.gov (Primary)
References for CVE-2020-12041
-
https://www.us-cert.gov/ics/advisories/icsma-20-170-04
Baxter Sigma Spectrum Infusion Pumps (Update A) | CISAThird Party Advisory;US Government Resource
Jump to