Vulnerability Details : CVE-2018-8868

Medtronic MyCareLink Patient Monitor, 24950 MyCareLink Monitor, all versions, and 24952 MyCareLink Monitor, all versions, contains debug code meant to test the functionality of the monitor's communication interfaces, including the interface between the monitor and implantable cardiac device. An attacker with physical access to the device can apply the other vulnerabilities within this advisory to access this debug functionality. This debug functionality provides the ability to read and write arbitrary memory values to implantable cardiac devices via inductive or short range wireless protocols. An attacker with close physical proximity to a target implantable cardiac device can use this debug functionality.
Publish Date : 2018-07-02 Last Update Date : 2019-10-09
Search Twitter   Search YouTube   Search Google

- CVSS Scores & Vulnerability Types

CVSS Score
6.9
Confidentiality Impact Complete (There is total information disclosure, resulting in all system files being revealed.)
Integrity Impact Complete (There is a total compromise of system integrity. There is a complete loss of system protection, resulting in the entire system being compromised.)
Availability Impact Complete (There is a total shutdown of the affected resource. The attacker can render the resource completely unavailable.)
Access Complexity Medium (The access conditions are somewhat specialized. Some preconditions must be satistified to exploit)
Authentication Not required (Authentication is not required to exploit the vulnerability.)
Gained Access None
Vulnerability Type(s)
CWE ID CWE id is not defined for this vulnerability

- Products Affected By CVE-2018-8868

# Product Type Vendor Product Version Update Edition Language
1 OS Medtronic 24950 Mycarelink Monitor Firmware - Version Details Vulnerabilities
2 OS Medtronic 24952 Mycarelink Monitor Firmware - Version Details Vulnerabilities

- Number Of Affected Versions By Product

Vendor Product Vulnerable Versions
Medtronic 24950 Mycarelink Monitor Firmware 1
Medtronic 24952 Mycarelink Monitor Firmware 1

- References For CVE-2018-8868

https://ics-cert.us-cert.gov/advisories/ICSMA-18-179-01

- Vulnerability Conditions

Vulnerability is valid if product versions listed below are used TOGETHER WITH(AND)
Vulnerability is valid if product versions listed below are used TOGETHER WITH(AND)

- Metasploit Modules Related To CVE-2018-8868

There are not any metasploit modules related to this CVE entry (Please visit www.metasploit.com for more information)


CVE is a registred trademark of the MITRE Corporation and the authoritative source of CVE content is MITRE's CVE web site. CWE is a registred trademark of the MITRE Corporation and the authoritative source of CWE content is MITRE's CWE web site. OVAL is a registered trademark of The MITRE Corporation and the authoritative source of OVAL content is MITRE's OVAL web site.
Use of this information constitutes acceptance for use in an AS IS condition. There are NO warranties, implied or otherwise, with regard to this information or its use. Any use of this information is at the user's risk. It is the responsibility of user to evaluate the accuracy, completeness or usefulness of any information, opinion, advice or other content. EACH USER WILL BE SOLELY RESPONSIBLE FOR ANY consequences of his or her direct or indirect use of this web site. ALL WARRANTIES OF ANY KIND ARE EXPRESSLY DISCLAIMED. This site will NOT BE LIABLE FOR ANY DIRECT, INDIRECT or any other kind of loss.