Medtronic MMT 508 MiniMed insulin pump, 522 / MMT - 722 Paradigm REAL-TIME, 523 / MMT - 723 Paradigm Revel, 523K / MMT - 723K Paradigm Revel, and 551 / MMT - 751 MiniMed 530G The models identified above, when paired with a remote controller and having the "easy bolus" and "remote bolus" options enabled (non-default), are vulnerable to a capture-replay attack. An attacker can capture the wireless transmissions between the remote controller and the pump and replay them to cause an insulin (bolus) delivery.
Published 2018-08-13 21:48:01
Updated 2019-10-09 23:35:12
Source ICS-CERT
View at NVD,   CVE.org
Vulnerability category: BypassGain privilege

Products affected by CVE-2018-14781

Exploit prediction scoring system (EPSS) score for CVE-2018-14781

0.07%
Probability of exploitation activity in the next 30 days EPSS Score History
~ 31 %
Percentile, the proportion of vulnerabilities that are scored at or less

CVSS scores for CVE-2018-14781

Base Score Base Severity CVSS Vector Exploitability Score Impact Score Score Source First Seen
2.9
LOW AV:A/AC:M/Au:N/C:N/I:P/A:N
5.5
2.9
NIST
5.3
MEDIUM CVSS:3.0/AV:A/AC:H/PR:N/UI:N/S:U/C:N/I:H/A:N
1.6
3.6
NIST

CWE ids for CVE-2018-14781

  • When an actor claims to have a given identity, the product does not prove or insufficiently proves that the claim is correct.
    Assigned by: nvd@nist.gov (Primary)
  • A capture-replay flaw exists when the design of the product makes it possible for a malicious user to sniff network traffic and bypass authentication by replaying it to the server in question to the same effect as the original message (or with minor changes).
    Assigned by: ics-cert@hq.dhs.gov (Secondary)

References for CVE-2018-14781