Medtronic N'Vision Clinician Programmer 8840 N'Vision Clinician Programmer, all versions, and 8870 N'Vision removable Application Card, all versions. The 8840 Clinician Programmer executes the application program from the 8870 Application Card. An attacker with physical access to an 8870 Application Card and sufficient technical capability can modify the contents of this card, including the binary executables. If modified to bypass protection mechanisms, this malicious code will be run when the card is inserted into an 8840 Clinician Programmer.
Published 2018-07-13 19:29:00
Updated 2019-10-09 23:32:58
Source ICS-CERT
View at NVD,   CVE.org

Exploit prediction scoring system (EPSS) score for CVE-2018-10631

0.07%
Probability of exploitation activity in the next 30 days EPSS Score History
~ 31 %
Percentile, the proportion of vulnerabilities that are scored at or less

CVSS scores for CVE-2018-10631

Base Score Base Severity CVSS Vector Exploitability Score Impact Score Score Source First Seen
4.6
MEDIUM AV:L/AC:L/Au:N/C:P/I:P/A:P
3.9
6.4
NIST
6.8
MEDIUM CVSS:3.0/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
0.9
5.9
NIST

CWE ids for CVE-2018-10631

  • The product does not use or incorrectly uses a protection mechanism that provides sufficient defense against directed attacks against the product.
    Assigned by:
    • ics-cert@hq.dhs.gov (Secondary)
    • nvd@nist.gov (Primary)

References for CVE-2018-10631

Products affected by CVE-2018-10631