Vulnerability Details : CVE-2017-9657
Under specific 802.11 network conditions, a partial re-association of the Philips IntelliVue MX40 Version B.06.18 WLAN monitor to the central monitoring station is possible. In this state, the central monitoring station can indicate the MX40 is not connected or associated to the central monitor, and thus should be operating in local monitoring mode (local audio-on, screen-on), but the MX40 WLAN itself can instead still be operating in telemetry mode (local audio-off, screen-off). If a patient experiences an alarm event and clinical staff expects the MX40 to provide local alarming when it is not available from the local device, a delay of treatment can occur. CVSS v3 base score: 6.5, CVSS vector string: AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H. Philips has released software update, Version B.06.18, to fix the improper cleanup on thrown exception vulnerability, and implement mitigations to reduce the risk associated with the improper handling of exceptional conditions vulnerability. The software update implements messaging and alarming on the MX40 and at the central monitoring station, when the MX40 disconnects from the access point.
Products affected by CVE-2017-9657
- cpe:2.3:o:philips:intellivue_mx40_firmware:*:*:*:*:*:*:*:*
Exploit prediction scoring system (EPSS) score for CVE-2017-9657
0.49%
Probability of exploitation activity in the next 30 days
EPSS Score History
~ 63 %
Percentile, the proportion of vulnerabilities that are scored at or less
CVSS scores for CVE-2017-9657
Base Score | Base Severity | CVSS Vector | Exploitability Score | Impact Score | Score Source | First Seen |
---|---|---|---|---|---|---|
3.3
|
LOW | AV:A/AC:L/Au:N/C:N/I:N/A:P |
6.5
|
2.9
|
NIST | |
6.5
|
MEDIUM | CVSS:3.0/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H |
2.8
|
3.6
|
NIST |
CWE ids for CVE-2017-9657
-
The product does not clean up its state or incorrectly cleans up its state when an exception is thrown, leading to unexpected state or control flow.Assigned by: ics-cert@hq.dhs.gov (Secondary)
-
The product does not handle or incorrectly handles an exceptional condition.Assigned by: nvd@nist.gov (Primary)
References for CVE-2017-9657
-
https://ics-cert.us-cert.gov/advisories/ICSMA-17-255-01
Philips' IntelliView MX40 Patient Worn Monitor (WLAN) Vulnerabilities | CISAThird Party Advisory;US Government Resource
-
https://www.usa.philips.com/healthcare/about/customer-support/product-security
Product Security | PhilipsVendor Advisory
-
http://www.securityfocus.com/bid/100813
Philips IntelliView MX40 Patient Worn Monitor Multiple Denial of Service VulnerabilitiesVDB Entry;Third Party Advisory