An issue was discovered in St. Jude Medical Merlin@home, versions prior to Version 8.2.2 (RF models: EX1150; Inductive models: EX1100; and Inductive models: EX1100 with MerlinOnDemand capability). The identities of the endpoints for the communication channel between the transmitter and St. Jude Medical's web site, Merlin.net, are not verified. This may allow a man-in-the-middle attacker to access or influence communications between the identified endpoints.
Published 2017-02-13 22:59:00
Updated 2023-06-26 19:38:41
Source ICS-CERT
View at NVD,   CVE.org

Products affected by CVE-2017-5149

Exploit prediction scoring system (EPSS) score for CVE-2017-5149

0.08%
Probability of exploitation activity in the next 30 days EPSS Score History
~ 33 %
Percentile, the proportion of vulnerabilities that are scored at or less

CVSS scores for CVE-2017-5149

Base Score Base Severity CVSS Vector Exploitability Score Impact Score Score Source First Seen
6.8
MEDIUM AV:N/AC:M/Au:N/C:P/I:P/A:P
8.6
6.4
NIST
8.9
HIGH CVSS:3.0/AV:N/AC:H/PR:N/UI:N/S:C/C:L/I:H/A:H
2.2
6.0
NIST