Abbott Laboratories pacemakers manufactured prior to Aug 28, 2017 do not restrict or limit the number of correctly formatted "RF wake-up" commands that can be received, which may allow a nearby attacker to repeatedly send commands to reduce pacemaker battery life. CVSS v3 base score: 5.3, CVSS vector string: AV:A/AC:H/PR:N/UI:N/S:U/C:N/I:N/A:H. Abbott has developed a firmware update to help mitigate the identified vulnerabilities.
Published 2018-04-25 13:29:00
Updated 2019-10-09 23:23:11
Source ICS-CERT
View at NVD,   CVE.org

Products affected by CVE-2017-12714

Exploit prediction scoring system (EPSS) score for CVE-2017-12714

0.05%
Probability of exploitation activity in the next 30 days EPSS Score History
~ 19 %
Percentile, the proportion of vulnerabilities that are scored at or less

CVSS scores for CVE-2017-12714

Base Score Base Severity CVSS Vector Exploitability Score Impact Score Score Source First Seen
6.1
MEDIUM AV:A/AC:L/Au:N/C:N/I:N/A:C
6.5
6.9
NIST
6.5
MEDIUM CVSS:3.0/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H
2.8
3.6
NIST

CWE ids for CVE-2017-12714

  • The product operates in an environment in which power is a limited resource that cannot be automatically replenished, but the product does not properly restrict the amount of power that its operation consumes.
    Assigned by:
    • ics-cert@hq.dhs.gov (Secondary)
    • nvd@nist.gov (Primary)

References for CVE-2017-12714

Jump to
This web site uses cookies for managing your session, storing preferences, website analytics and additional purposes described in our privacy policy.
By using this web site you are agreeing to CVEdetails.com terms of use!