Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior accept drug libraries, firmware updates, pump commands, and unauthorized configuration changes from unauthenticated devices on the host network. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.
Published 2019-03-25 18:29:00
Updated 2019-10-09 23:14:06
Source ICS-CERT
View at NVD,   CVE.org

Products affected by CVE-2015-3956

Exploit prediction scoring system (EPSS) score for CVE-2015-3956

0.21%
Probability of exploitation activity in the next 30 days EPSS Score History
~ 58 %
Percentile, the proportion of vulnerabilities that are scored at or less

CVSS scores for CVE-2015-3956

Base Score Base Severity CVSS Vector Exploitability Score Impact Score Score Source First Seen
10.0
HIGH AV:N/AC:L/Au:N/C:C/I:C/A:C
10.0
10.0
NIST
9.8
CRITICAL CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
3.9
5.9
NIST

CWE ids for CVE-2015-3956

References for CVE-2015-3956