Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior give unauthenticated users root privileges on Port 23/TELNET by default. An unauthorized user could issue commands to the pump. Hospira recommends that customers close Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.
Published 2019-03-25 17:29:01
Updated 2019-10-09 23:14:06
Source ICS-CERT
View at NVD,   CVE.org
Vulnerability category: BypassGain privilege

Exploit prediction scoring system (EPSS) score for CVE-2015-3954

Probability of exploitation activity in the next 30 days: 0.21%

Percentile, the proportion of vulnerabilities that are scored at or less: ~ 58 % EPSS Score History EPSS FAQ

CVSS scores for CVE-2015-3954

Base Score Base Severity CVSS Vector Exploitability Score Impact Score Score Source
10.0
HIGH AV:N/AC:L/Au:N/C:C/I:C/A:C
10.0
10.0
NIST
9.8
CRITICAL CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
3.9
5.9
NIST

CWE ids for CVE-2015-3954

  • The product does not perform or incorrectly performs an authorization check when an actor attempts to access a resource or perform an action.
    Assigned by:
    • ics-cert@hq.dhs.gov (Secondary)
    • nvd@nist.gov (Primary)

References for CVE-2015-3954

Products affected by CVE-2015-3954