CVEdetails.com the ultimate security vulnerability data source
(e.g.: CVE-2009-1234 or 2010-1234 or 20101234)
Log In   Register
Vulnerability Feeds & WidgetsNew   www.itsecdb.com  

Vulnerability Details : CVE-2015-3954

Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior give unauthenticated users root privileges on Port 23/TELNET by default. An unauthorized user could issue commands to the pump. Hospira recommends that customers close Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.
Publish Date : 2019-03-25 Last Update Date : 2019-10-09
Search Twitter   Search YouTube   Search Google

- CVSS Scores & Vulnerability Types

CVSS Score
10.0
Confidentiality Impact Complete (There is total information disclosure, resulting in all system files being revealed.)
Integrity Impact Complete (There is a total compromise of system integrity. There is a complete loss of system protection, resulting in the entire system being compromised.)
Availability Impact Complete (There is a total shutdown of the affected resource. The attacker can render the resource completely unavailable.)
Access Complexity Low (Specialized access conditions or extenuating circumstances do not exist. Very little knowledge or skill is required to exploit. )
Authentication Not required (Authentication is not required to exploit the vulnerability.)
Gained Access None
Vulnerability Type(s)
CWE ID 285

- Products Affected By CVE-2015-3954

# Product Type Vendor Product Version Update Edition Language
1 OS Pifzer Plum A+ Infusion System Firmware 13.4 Version Details Vulnerabilities
2 OS Pifzer Plum A+3 Infusion System Firmware 13.6 Version Details Vulnerabilities
3 OS Pifzer Symbiq Infusion System Firmware 3.13 Version Details Vulnerabilities

- Number Of Affected Versions By Product

Vendor Product Vulnerable Versions
Pifzer Plum A+ Infusion System Firmware 1
Pifzer Plum A+3 Infusion System Firmware 1
Pifzer Symbiq Infusion System Firmware 1

- References For CVE-2015-3954

https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01

- Vulnerability Conditions

Vulnerability is valid if product versions listed below are used TOGETHER WITH(AND)
Vulnerability is valid if product versions listed below are used TOGETHER WITH(AND)
Vulnerability is valid if product versions listed below are used TOGETHER WITH(AND)

- Metasploit Modules Related To CVE-2015-3954

There are not any metasploit modules related to this CVE entry (Please visit www.metasploit.com for more information)


CVE is a registred trademark of the MITRE Corporation and the authoritative source of CVE content is MITRE's CVE web site. CWE is a registred trademark of the MITRE Corporation and the authoritative source of CWE content is MITRE's CWE web site. OVAL is a registered trademark of The MITRE Corporation and the authoritative source of OVAL content is MITRE's OVAL web site.
Use of this information constitutes acceptance for use in an AS IS condition. There are NO warranties, implied or otherwise, with regard to this information or its use. Any use of this information is at the user's risk. It is the responsibility of user to evaluate the accuracy, completeness or usefulness of any information, opinion, advice or other content. EACH USER WILL BE SOLELY RESPONSIBLE FOR ANY consequences of his or her direct or indirect use of this web site. ALL WARRANTIES OF ANY KIND ARE EXPRESSLY DISCLAIMED. This site will NOT BE LIABLE FOR ANY DIRECT, INDIRECT or any other kind of loss.