Wireless keys are stored in plain text on Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.
Published 2019-03-25 16:29:00
Updated 2019-10-09 23:14:05
Source ICS-CERT
View at NVD,   CVE.org
Vulnerability category: Information leak

Exploit prediction scoring system (EPSS) score for CVE-2015-3952

0.16%
Probability of exploitation activity in the next 30 days EPSS Score History
~ 52 %
Percentile, the proportion of vulnerabilities that are scored at or less

CVSS scores for CVE-2015-3952

Base Score Base Severity CVSS Vector Exploitability Score Impact Score Score Source First Seen
5.0
MEDIUM AV:N/AC:L/Au:N/C:P/I:N/A:N
10.0
2.9
NIST
7.5
HIGH CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N
3.9
3.6
NIST

CWE ids for CVE-2015-3952

References for CVE-2015-3952

Products affected by CVE-2015-3952

This web site uses cookies for managing your session, storing preferences, website analytics and additional purposes described in our privacy policy.
By using this web site you are agreeing to CVEdetails.com terms of use!