Vulnerability Details : CVE-2015-3952
Wireless keys are stored in plain text on Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.
Vulnerability category: Information leak
Exploit prediction scoring system (EPSS) score for CVE-2015-3952
0.16%
Probability of exploitation activity in the next 30 days
EPSS Score History
~ 52 %
Percentile, the proportion of vulnerabilities that are scored at or less