Vulnerability Details : CVE-2014-5433
An unauthenticated remote attacker may be able to execute commands to view wireless account credentials that are stored in cleartext on Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16, which may allow an attacker to gain access the host network. Baxter has released a new version of the SIGMA Spectrum Infusion System, Version 8, which incorporates hardware and software changes.
Exploit prediction scoring system (EPSS) score for CVE-2014-5433
Probability of exploitation activity in the next 30 days: 0.44%
Percentile, the proportion of vulnerabilities that are scored at or less: ~ 72 % EPSS Score History EPSS FAQ
CVSS scores for CVE-2014-5433
Base Score | Base Severity | CVSS Vector | Exploitability Score | Impact Score | Score Source |
---|---|---|---|---|---|
7.5
|
HIGH | AV:N/AC:L/Au:N/C:P/I:P/A:P |
10.0
|
6.4
|
NIST |
9.8
|
CRITICAL | CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H |
3.9
|
5.9
|
NIST |
CWE ids for CVE-2014-5433
-
Assigned by: nvd@nist.gov (Primary)
-
The product stores sensitive information in cleartext within a resource that might be accessible to another control sphere.Assigned by: ics-cert@hq.dhs.gov (Secondary)
References for CVE-2014-5433
-
https://ics-cert.us-cert.gov/advisories/ICSA-15-181-01
Baxter SIGMA Spectrum Infusion System Vulnerabilities | CISAMitigation;US Government Resource;Third Party Advisory
Products affected by CVE-2014-5433
- cpe:2.3:o:baxter:sigma_spectrum_infusion_system_firmware:6.05:*:*:*:*:*:*:*