An unauthenticated remote attacker may be able to execute commands to view wireless account credentials that are stored in cleartext on Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16, which may allow an attacker to gain access the host network. Baxter has released a new version of the SIGMA Spectrum Infusion System, Version 8, which incorporates hardware and software changes.
Published 2019-03-26 16:29:00
Updated 2019-10-09 23:11:11
Source ICS-CERT
View at NVD,   CVE.org

Exploit prediction scoring system (EPSS) score for CVE-2014-5433

Probability of exploitation activity in the next 30 days: 0.44%

Percentile, the proportion of vulnerabilities that are scored at or less: ~ 72 % EPSS Score History EPSS FAQ

CVSS scores for CVE-2014-5433

Base Score Base Severity CVSS Vector Exploitability Score Impact Score Score Source
7.5
HIGH AV:N/AC:L/Au:N/C:P/I:P/A:P
10.0
6.4
NIST
9.8
CRITICAL CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
3.9
5.9
NIST

CWE ids for CVE-2014-5433

References for CVE-2014-5433

Products affected by CVE-2014-5433