Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 is remotely accessible via Port 22/SSH without authentication. A remote attacker may be able to make unauthorized configuration changes to the WBM, as well as issue commands to access account credentials and shared keys. Baxter asserts that this vulnerability only allows access to features and functionality on the WBM and that the SIGMA Spectrum infusion pump cannot be controlled from the WBM. Baxter has released a new version of the SIGMA Spectrum Infusion System, Version 8, which incorporates hardware and software changes.
Published 2019-03-26 16:29:00
Updated 2019-10-09 23:11:11
Source ICS-CERT
View at NVD,   CVE.org
Vulnerability category: BypassGain privilege

Exploit prediction scoring system (EPSS) score for CVE-2014-5432

0.36%
Probability of exploitation activity in the next 30 days EPSS Score History
~ 69 %
Percentile, the proportion of vulnerabilities that are scored at or less

CVSS scores for CVE-2014-5432

Base Score Base Severity CVSS Vector Exploitability Score Impact Score Score Source First Seen
7.5
HIGH AV:N/AC:L/Au:N/C:P/I:P/A:P
10.0
6.4
NIST
9.8
CRITICAL CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
3.9
5.9
NIST

CWE ids for CVE-2014-5432

  • When an actor claims to have a given identity, the product does not prove or insufficiently proves that the claim is correct.
    Assigned by: nvd@nist.gov (Primary)
  • Assigned by: ics-cert@hq.dhs.gov (Secondary)

References for CVE-2014-5432

Products affected by CVE-2014-5432

This web site uses cookies for managing your session, storing preferences, website analytics and additional purposes described in our privacy policy.
By using this web site you are agreeing to CVEdetails.com terms of use!