Vulnerability Details : CVE-2014-5406
The Hospira LifeCare PCA Infusion System before 7.0 does not validate network traffic associated with sending a (1) drug library, (2) software update, or (3) configuration change, which allows remote attackers to modify settings or medication data via packets on the (a) TELNET, (b) HTTP, (c) HTTPS, or (d) UPNP port. NOTE: this issue might overlap CVE-2015-3459.
Products affected by CVE-2014-5406
- cpe:2.3:o:hospira:lifecare_pcainfusion_firmware:*:*:*:*:*:*:*:*
Exploit prediction scoring system (EPSS) score for CVE-2014-5406
0.51%
Probability of exploitation activity in the next 30 days
EPSS Score History
~ 76 %
Percentile, the proportion of vulnerabilities that are scored at or less
CVSS scores for CVE-2014-5406
Base Score | Base Severity | CVSS Vector | Exploitability Score | Impact Score | Score Source | First Seen |
---|---|---|---|---|---|---|
9.3
|
HIGH | AV:N/AC:M/Au:N/C:C/I:C/A:C |
8.6
|
10.0
|
NIST |
CWE ids for CVE-2014-5406
-
The product does not sufficiently verify the origin or authenticity of data, in a way that causes it to accept invalid data.Assigned by: nvd@nist.gov (Primary)
References for CVE-2014-5406
-
http://www.fda.gov/MedicalDevices/Safety/AlertsandNotices/ucm446809.htm
Page Not Found | FDAThird Party Advisory;US Government Resource
-
https://xs-sniper.com/blog/2015/06/08/hospira-plum-a-infusion-pump-vulnerabilities/
-
https://ics-cert.us-cert.gov/advisories/ICSA-15-125-01
Access Denied | CISAThird Party Advisory;US Government Resource
Jump to